Chapter 2. Laboratory Techniques

2.1 Safety in the Lab

Types of Pipettes

Mechanical Pipettes

Mechanical pipettes, or micropipettes, are used for very small volume transfers (i.e., less than 1 mL) where a high degree of precision is required.

Figure 1-1. Mechanical or micropipette.

Micropipettes come in several size ranges. Depending on the size, different color-coded disposable tips are used. The three most common size ranges are 100 to 1000 μL (0.1 mL to 1 mL), which take blue tips, and 10 to 100 μL (0.01 mL to 0.1 mL) and 1 to 20 μL (0.001 mL to 0.02 mL), both of which take yellow tips. This semester you will be using the 10 to 100 μL size. The volume to be transferred is set by turning the volume adjustment rings. A disposable tip fits on the end of the pipette. By depressing the plunger to the first stop point and placing the tip in the liquid and slowly releasing the plunger, the exact amount to be transferred can be drawn up into the tip. To expel this volume, depress the plunger to the second stop point. In some cases, micropipettes are fitted with a tip ejector to remove the tip.

Volumetric Pipettes

Volumetric, or transfer, pipettes are designed to deliver a single precise volume. The volume of the pipette is indicated near the top of the pipette (i.e., 5 mL). These pipettes do not have gradation markings and only have a single mark or engraved ring above the bulb in the pipette indicating the volume the pipette is designed to contain.

Figure 1-2. Volumetric pipette.

Fluid must be drawn up in the pipette to above the mark or engraved ring and then released slowly until the bottom of the meniscus of the liquid is exactly at the engraved ring or mark (the tip of the pipette should be touching the wall of the sample vessel as fluid is released). To transfer this volume to a second container, touch the pipette tip to the inside wall of the new container and allow the liquid to drain out.

Graduated Pipettes

There are two general types of graduated pipettes: Mohr and serological. Each of these types comes in various sizes which are used for various volumes. You should try to use the pipette that is just big enough for the volume you are going to transfer. For example, if the volume you need to transfer is 1.5 mL, the best pipette to accomplish this with the highest accuracy would be a 2 mL pipette (not a 1 mL or a 5 mL or larger).

Mohr pipettes are designed to transfer a volume by moving the meniscus of the liquid between two gradation marks on the pipette. The gradation markings on the Mohr pipette do not go all the way to the tip of the pipette but stop before the pipette begins to narrow (see Figure 1-3).

Figure 1-3. A comparison of the tips of a serological pipette (left) and a Mohr pipette (right).

To accurately transfer fluid with a Mohr pipette, the meniscus must be precisely on a calibration mark both at the beginning and at the end of a transfer. Near the top of this type of pipette you will find the total volume indicated and the size of the smallest gradations (i.e., 10 mL in 1/10, the total volume of the pipette is 10 mL and it is graduated in 1/10 mL increments). The marking near the tip of the pipette should be the same as the total volume listed near the top of the Mohr pipette.

Serological pipettes are graduated to deliver (there is no base mark). The appropriate amount of fluid is drawn into the pipette (with the meniscus precisely on the correct mark) and the entire amount is transferred. Since serological pipettes are labeled with the zero mark at the top of the pipette, you will need to subtract the amount you are going to pipette from the total volume of the pipette to determine the exact mark to fill the pipette to. For example, if you are going to pipette 6.5 mL using a 10 mL pipette you would fill the pipette to the 3.5 mL mark. To determine the total volume of the pipette, look near the top of this type of pipette. There you will find the total volume indicated and the size of the smallest gradations. There are two types of serological pipettes. Those with a single painted or frosted ring at the top should be allowed to simply drain with the tip placed against the side of the receiving vessel. Those with double rings are designed to be “blown out” by pushing a small amount of air out of the pipette, completely emptying it.

ALWAYS USE PI-PUMPS FOR FILLING PIPETTES.
Figure 1-4. Pi-pump.

A pi-pump is a device designed to safely and easily draw fluid up into a pipette. Since not all pipettes have the same size diameter, there are two different sizes of pi-pumps made to fit the various sizes of pipettes: blue for pipettes up to 2 mL and green for 5 or 10 mL pipettes. The pi-pump has a neoprene gasket with a hole in it designed to fit over the top end of a pipette. The pi-pump has a neoprene gasket with a hole in it designed to fit over the top end of a pipette. To draw up fluid into the pipette, operate the wheel control on the pi-pump. To expel the fluid from the pipette you can either depress the plunger on the top of the pi-pump or for more control rotate the wheel control in the opposite direction.